Validation of 3-D freehand ultrasound for the determination of the hip joint centre
Peters, A, Baker, RJ and Sangeux, M 2010, 'Validation of 3-D freehand ultrasound for the determination of the hip joint centre' , Gait & Posture, 31 (4) , pp. 530-532.
| PDF - Published Version Restricted to Repository staff only Download (145kB) | Request a copy |
Abstract
This paper describes the calibration process for 3-D free hand ultrasound (3-DUS). The Cambridge calibration technique was determined to produce the most repeatable and accurate results when determining the location of a reference object within a water bath. This note additionally validates 3-DUS as a potential measurement tool for the determination of the hip joint centre. It shows that 3-DUS can accurately determine the inter-HJC distance to within 4 � 2 mm. This indicates potential for 3-DUS to be used as a gold standard measurement tool in the identification of the hip joint centre
| Item Type: | Article |
|---|---|
| Themes: | Health and Wellbeing |
| Schools: | Colleges and Schools > College of Health & Social Care Colleges and Schools > College of Health & Social Care > School of Health Sciences |
| Journal or Publication Title: | Gait & Posture |
| Publisher: | Elsevier |
| Refereed: | Yes |
| ISSN: | 0966-6362 |
| Depositing User: | RH Shuttleworth |
| Date Deposited: | 11 May 2011 10:31 |
| Last Modified: | 11 May 2011 10:32 |
| URI: | http://usir.salford.ac.uk/id/eprint/15742 |