Skip to the content

Pressure and breast thickness in mammography an exploratory calibration study

Hogg, P, Taylor, M, Szczepura, K, Mercer, C and Denton, E 2013, 'Pressure and breast thickness in mammography an exploratory calibration study' , British journal of Radiology, 86 (2) , pp. 1-7.

[img] PDF
Restricted to Repository staff only

Download (326kB) | Request a copy

Abstract

Objective To perform a calibration study to provide data to help improve consistency in the pressure that is applied during mammography. Methods Automatic readouts of breast thickness accuracy vary between mammography machines; therefore, one machine was selected for calibration. 250 randomly selected patients were invited to participate; 235 agreed, and 940 compression data sets were recorded (breast thickness, breast density and pressure). Pressure (measured in decanewtons) was increased from 5 daN through 1-daN intervals until the practitioner felt that the pressure was appropriate for imaging; at each pressure increment, breast thickness was recorded. Results Graphs were generated and equations derived; second-order polynomial trend lines were applied using the method of least squares. No difference existed between breast densities, but a difference did exist between “small” (15×29 cm) and “medium/large” (18×24/24×30 cm) paddles. Accordingly, data were combined. Graphs show changes in thickness from 5-daN pressure for craniocaudal and mediolateral oblique views for the small and medium/large paddles combined. Graphs were colour coded into three segments indicating high, intermediate and low gradients [≤−2 (light grey); −1.99 to −1 (mid-grey); and ≥−0.99 (dark grey)]. We propose that 13 daN could be an appropriate termination pressure on this mammography machine. Conclusion Using patient compression data we have calibrated a mammography machine to determine its breast compression characteristics. This calibration data could be used to guide practice to minimise pressure variations between practitioners, thereby improving patient experience and reducing potential variation in image quality. Advances in knowledge For the first time, pressure–thickness graphs are now available to help guide mammographers in the application of pressure.

Item Type: Article
Themes: Health and Wellbeing
Schools: Schools > School of Health Sciences > Centre for Health Sciences Research
Schools > School of Health Sciences
Journal or Publication Title: British journal of Radiology
Publisher: British Institute of Radiology
Refereed: Yes
ISSN: 0007-1285
Related URLs:
Funders: Non funded research
Depositing User: P Hogg
Date Deposited: 08 Oct 2013 11:16
Last Modified: 30 Nov 2015 23:43
URI: http://usir.salford.ac.uk/id/eprint/29530

Actions (login required)

Edit record (repository staff only) Edit record (repository staff only)

Downloads

Downloads per month over past year