Validation of 3-D freehand ultrasound for the determination of the hip joint centre

Peters, A, Baker, RJ ORCID: https://orcid.org/0000-0003-4759-4216 and Sangeux, M 2010, 'Validation of 3-D freehand ultrasound for the determination of the hip joint centre' , Gait & Posture, 31 (4) , pp. 530-532.

[img] PDF - Published Version
Restricted to Repository staff only

Download (149kB) | Request a copy

Abstract

This paper describes the calibration process for 3-D free hand ultrasound (3-DUS). The Cambridge calibration technique was determined to produce the most repeatable and accurate results when determining the location of a reference object within a water bath. This note additionally validates 3-DUS as a potential measurement tool for the determination of the hip joint centre. It shows that 3-DUS can accurately determine the inter-HJC distance to within 4 � 2 mm. This indicates potential for 3-DUS to be used as a gold standard measurement tool in the identification of the hip joint centre

Item Type: Article
Themes: Health and Wellbeing
Schools: Schools > School of Health Sciences
Journal or Publication Title: Gait & Posture
Publisher: Elsevier
Refereed: Yes
ISSN: 0966-6362
Depositing User: RH Shuttleworth
Date Deposited: 11 May 2011 09:31
Last Modified: 16 Feb 2022 11:23
URI: https://usir.salford.ac.uk/id/eprint/15742

Actions (login required)

Edit record (repository staff only) Edit record (repository staff only)

Downloads

Downloads per month over past year