Analysis of effective and organ dose estimation in CT when using mA modulation : a single scanner pilot study

Tootell, AK ORCID: https://orcid.org/0000-0002-8567-8659, Szczepura, K ORCID: https://orcid.org/0000-0002-2566-3308 and Hogg, P ORCID: https://orcid.org/0000-0002-6506-0827 2017, 'Analysis of effective and organ dose estimation in CT when using mA modulation : a single scanner pilot study' , Radiography, 23 (2) , pp. 159-166.

[img]
Preview
PDF - Accepted Version
Download (439kB) | Preview

Abstract

Effective dose (ED) estimation in CT examinations can be obtained by combining dose length product (DLP) with published ED per DLP coefficients or performed using software. These methods do not account for tube current (mA) modulation which is influenced by patient size.
Aim
To compare different methods of organ and ED estimation to measured values when using mA modulation in CT chest, abdomen and pelvis examinations.
Method
Organ doses from CT of the chest, abdomen and pelvis were measured using digital dosimeters and a dosimetry phantom. ED was calculated. Six methods of estimating ED accounting for mA modulation were performed using ImPACT CTDosimetry and Dose Length Product to ED coefficients. Corrections for the phantom mass were applied resulting in 12 estimation methods. Estimated organ doses from ImPACT CTDosimtery were compared to measured values.
Results
Calculated EDs were; chest 12.35 mSv (±1.48 mSv); abdomen 8.74 mSv (±1.36 mSv) and pelvis 4.68 mSv (±0.75 mSv). There was over estimation in all three anatomical regions. Correcting for phantom mass improved agreement between measured and estimated ED. Organ doses showed overestimation of dose inside the scan range and underestimation outside the scan range.
Conclusion
Reasonable estimation of effective dose for CT of the chest and abdomen can be obtained using ImPACT CTDosimetry software or k-coefficients. Further work is required to improve the accuracy of ED estimation from CT of the pelvis. Accuracy of organ dose estimation has been shown to depend on the inclusion or exclusion of the organ from the scan range.

Item Type: Article
Schools: Schools > School of Health and Society > Centre for Health Sciences Research
Journal or Publication Title: Radiography
Publisher: Elsevier
ISSN: 1078-8174
Related URLs:
Depositing User: AK Tootell
Date Deposited: 27 Feb 2017 13:22
Last Modified: 15 Feb 2022 21:45
URI: https://usir.salford.ac.uk/id/eprint/41456

Actions (login required)

Edit record (repository staff only) Edit record (repository staff only)

Downloads

Downloads per month over past year