Szczepura, K ORCID: https://orcid.org/0000-0002-2566-3308, Tomkinson, D and Manning, D
2017,
'Impact of tube current modulation on lesion conspicuity index in hi-resolution chest computed tomography'
, Proceedings of SPIE Volume 10136, Medical Imaging 2017: Image Perception, Observer Performance, and Technology Assessment, 101361A, 10136
.
![]() |
PDF
- Accepted Version
Restricted to Repository staff only Download (195kB) | Request a copy |
![]() |
Microsoft Word
- Accepted Version
Restricted to Repository staff only Download (2MB) | Request a copy |
Abstract
Tube current modulation is a method employed in the use of CT in an attempt to optimize radiation dose to the patient. The acceptable noise (noise index) can be varied, based on the level of optimization required; higher accepted noise reduces the patient dose. Recent research [1] suggests that measuring the conspicuity index (C.I.) of focal lesions within an image is more reflective of a clinical reader's ability to perceive focal lesions than traditional physical measures such as contrast to noise (CNR) and signal to noise ratio (SNR). Software has been developed and validated to calculate the C.I. in DICOM images. The aim of this work is assess the impact of tube current modulation on conspicuity index and CTDIvol, to indicate the benefits and limitations of tube current modulation on lesion detectability.
Method: An anthropomorphic chest phantom was used “Lungman” with inserted lesions of varying size and HU (see table below) a range of Hounsfield units and sizes were used to represent the variation in lesion Hounsfield units found. This meant some lesions had negative Hounsfield unit values.
Item Type: | Article |
---|---|
Schools: | Schools > School of Health and Society > Centre for Health Sciences Research |
Journal or Publication Title: | Proceedings of SPIE Volume 10136, Medical Imaging 2017: Image Perception, Observer Performance, and Technology Assessment, 101361A |
Publisher: | SPIE Digital Library |
ISSN: | 0277-786X |
Depositing User: | K Szczepura |
Date Deposited: | 14 Mar 2019 14:25 |
Last Modified: | 16 Feb 2022 01:16 |
URI: | https://usir.salford.ac.uk/id/eprint/50422 |