Ali, MA, Hogg, P ORCID: https://orcid.org/0000-0002-6506-0827 and England, A
ORCID: https://orcid.org/0000-0001-6333-7776
2019,
'Dose optimisation in paediatric radiography - using regression models to investigate the relative impact of acquisition factors on image quality and radiation dose'
, Physica Medica : European Journal of Medical Physics, 68
, pp. 61-68.
|
PDF
- Published Version
Available under License Creative Commons Attribution Non-commercial No Derivatives 4.0. Download (1MB) | Preview |
|
![]() |
PDF
- Accepted Version
Restricted to Repository staff only Download (1MB) | Request a copy |
Abstract
Objective: To investigate the optimum pelvis X-ray acquisition factors for a 10-year-old child. Secondly, to evaluate the impact of each acquisition factor on image quality (IQ) and radiation dose.
Method: Images were acquired using a pelvis phantom and a range of acquisition parameters; e.g. tube potential, additional filtration and source-to-image distance (SID). Automatic exposure control (AEC) was used with two orientations (head towards/away from two outer chambers) and three different chamber selections. Visual IQ was evaluated using relative and absolute-VGA methods. Radiation doses were measured by placing a dosimeter on the anterior surface of the phantom. Regression analysis was used to determine optimum parameters.
Results: The optimised technique (178.8 µGy), with diagnostic IQ, was with 89kVp, 130 cm SID and with 1 mm Al + 0.1 mm Cu filtration. This technique was with the head towards the two outer AEC chambers. Regression analysis showed that SID had the lowest impact on IQ (β = 0.002 95% CI −0.001 to 0.005) and dose (β = −0.96 95% CI −0.40 to −1.53). The impact of filtration on dose (β = −76.24 95% CI −86.76 to −85.72) was higher than tube potential (β = −13.44 95% CI −14.34 to −12.53). The following impact ratios were higher on IQ than radiation dose: filtration/kVp; 11.28 times, filtration/SID; 7.01 times and kVp/SID; 0.62 times.
Conclusion: Optimised parameters were identified as 89 kVp, 130 cm SID and with 1 mm Al + 0.1 mm Cu additional filtration. Regression analysis demonstrated that filtration and tube potential had the greatest effect on radiation dose and IQ, respectively.
Item Type: | Article |
---|---|
Schools: | Schools > School of Health and Society > Centre for Health Sciences Research |
Journal or Publication Title: | Physica Medica : European Journal of Medical Physics |
Publisher: | Elsevier |
ISSN: | 1120-1797 |
Related URLs: | |
Depositing User: | USIR Admin |
Date Deposited: | 22 Jul 2019 07:52 |
Last Modified: | 16 Feb 2022 02:22 |
URI: | https://usir.salford.ac.uk/id/eprint/51850 |
Actions (login required)
![]() |
Edit record (repository staff only) |