Predictors of radiation dose for uterine artery embolisation are angiography system-dependent

Nocum, D, Robinson, J, Halaki, M, Båth, M, Thompson, JD ORCID: https://orcid.org/0000-0002-4786-7379, Thompson, N, Moscova, M, Liang, E, Mekiš, N and Reed, W 2022, 'Predictors of radiation dose for uterine artery embolisation are angiography system-dependent' , Journal of Radiological Protection, 42 (1) , 011502.

[img] PDF - Accepted Version
Restricted to Repository staff only until 17 January 2023.

Download (427kB) | Request a copy
[img] Microsoft Word - Accepted Version
Restricted to Repository staff only

Download (125kB) | Request a copy

Abstract

This study sought to achieve radiation dose reductions for patients receiving uterine artery embolisation (UAE) by evaluating radiation dose measurements for the preceding generation (Allura) and upgraded (Azurion) angiography system. Previous UAE regression models in the literature could not be applied to this centre’s practice due to being based on different angiography systems and radiation dose predictor variables. The aims of this study were to establish whether radiation dose is reduced with the upgraded angiography system and to develop a regression model to determine predictors of radiation dose specific to the upgraded angiography system. A comparison between Group I (Allura, n = 95) and Group II (Azurion, n = 95) demonstrated a significant reduction in KAP (kerma-area product) and Ka, r (reference air kerma) by 63% (143.2 Gy·cm2 vs 52.9 Gy·cm2; P < 0.001, d = 0.8) and 67% (0.6 Gy vs 0.2 Gy; P < 0.001, d = 0.8), respectively. The multivariable linear regression (MLR) model identified the UAE radiation dose predictors for KAP on the upgraded angiography system as total fluoroscopy dose, Ka, r, and total uterus volume. The predictive accuracy of the MLR model was assessed using a Bland-Altman plot. The mean difference was 0.39 Gy·cm2 and the limits of agreement (LoA) were +28.49 and -27.71 Gy·cm2, and thus illustrated no proportional bias. The resultant MLR model was considered system-dependent and validated the upgraded angiography system and its advance capabilities to significantly reduce radiation dose. Interventional radiologist and interventional radiographer familiarisation of the system’s features and the implementation of the newly established MLR model would further facilitate dose optimisation for all centres performing UAE procedures using the upgraded angiography system.

Item Type: Article
Schools: Schools > School of Health and Society > Centre for Health Sciences Research
Journal or Publication Title: Journal of Radiological Protection
Publisher: Institute of Physics
ISSN: 0952-4746
Related URLs:
Depositing User: JD Thompson
Date Deposited: 06 Jan 2022 14:23
Last Modified: 15 Feb 2022 16:51